Product Description
The model 1260A Impedance/Gain-Phase Analyzer is - without doubt - the most powerful, accurate and flexible Frequency Response Analyzer available today. In daily use by leading researchers wherever measurement integrity and experimental reliability are of paramount importance, the 1260A¡¯s solid reputation is frequently endorsed in published research papers in fields such as:
Corrosion studies
Battery research and fuel cells
Solar cells
LCD's
Bio-materials
Ceramics / composites
Electronic component development
Civil engineering
One of Solartron Analytical¡¯s extensive range of precision products designed to provide cost effective solutions for DC and AC analysis in electrochemical and materials research, the 1260A offers an outstanding measurement specification for impedance spectroscopy.
Wide frequency range
Spanning 10¦ÌHz to 32MHz with 0.015ppm resolution, the 1260A provides excellent coverage for virtually all chemical and molecular mechanisms - all in a single instrument.
Unbeatable accuracy
With an accuracy of 0.1%, 0.1o, measurements can be made with complete confidence, and even the most subtle changes in sample behavior detected and quantized
Noise free analysis
The 1260A uses Solartron Analytical¡¯s patented single-sine correlation technique, which inherently removes the noise and harmonic distortion which plagues lesser instruments.
Frequency resolution: 1 in 65 million (0.015ppm)
0.1%, 0.1o accuracy - unsurpassed by any similar instrument
Resolution to 0.001dB, 0.01o - capturing every detail
Measures impedances >100M?
2-, 3- and 4-terminal measurement configurations
Polarization voltage up to ¡À40.95V
Compatible with the renowned ZPlot software package to simplify experiments and optimize throughput.
Systems
When combined with other products from Solartron Analytical¡¯s range, including well-proven application software (ZPlot or SMaRT), the 1260A can form the heart of an advanced electrochemical and materials measurement system, to provide superb accuracy, flexibility and reliability - even for the most complex research problems.
Impedance measurement
Virtually every liquid and solid is able to pass current when a voltage is applied to it. If a variable (ac) voltage is applied to the material, the ratio of voltage to current is known as the impedance. The measured impedance varies with the frequency of the applied voltage in a way that is related to the properties of the liquid or solid. This may be due to the physical structure of the material, to chemical processes within it or a combination of both. The advantages of impedance measurement over other techniques include:-
Rapid acquisition of data
Accurate, repeatable measurements
Non-destructive
Highly adaptable to a wide variety of different applications
Ability to differentiate effects due to electrodes, diffusion, mass/charge transfer by analysis over different frequency ranges
Equivalent circuit/modeling
| Generator | |||
| Voltage mode | Current mode | ||
| AC Amplitude ¡Ü10MHz | 0 to 3 V rms | 0 to 60 mA rms | |
| >10MHz | 0 to 1 V rms | 0 to 20 mA rms | |
| Maximum AC resolution | 5 mV | 100 ¦ÌA | |
| DC Bias Range | ¡À40.95 V | ¡À100 mA | |
| Maximum DC resolution | 10 mV | 200 ¦ÌA | |
| Output impedance | 50 ?¡À1% | >200 k? at <1 kHz | |
| Frequency range | 10 ¦ÌHz to 32 MHz | ||
| Max resolution | 10 ¦ÌHz | ||
| Error: | ¡À100 ppm | ||
| Stability | 24 hrs ¡À1oC | ||
| ¡À10 ppm | |||
| Sweep types | Frequency (logarithmic or linear) | ||
| AC/DC voltage | |||
| AC/DC current | |||
| Maximum voltage hi to lo | ¡À46 V peak | ||
| Maximum voltage lo to ground | ¡À0.4 V peak | ||
| Maximum current | ¡À100 mA peak | ||
| Impedance lo to ground | 100 k?, <10 nF | ||
| Connector | Single BNC, floating shield | ||
| Output disable control | Contact closure or TTL logic 0 | ||
| Analyzers | |||
| 3 independent analyzers operating in parallel | |||
| Voltage mode (x2) | Current mode | ||
| Ranges | 30 mV, 300 mV, 3 V | 6 ¦ÌA, 60 ¦ÌA, 600 ¦ÌA, 6 mA, 60 mA | |
| Maximum resolution | 1 ¦ÌV | 200 pA | |
| Full scale peak | ¡À5 V | ¡À100 mA | |
| Inputs protected to | ¡À46 V | ¡À250 mA | |
| Connections | single/differential BNC | single BNC | |
| Shields | floating/grounded | - | |
| Coupling | DC or AC (-3dB at 1Hz) | DC or AC (-3dB at 1 Hz) | |
| Input impedance hi to shield | 1 M?, <35 pF | ¡Ý600 ¦ÌA range, 1 ? | |
| Input impedance shield to ground | 10 k?, 330 pF | <600 ¦ÌA range, 50 ? | |
| Limits of error | |||
| Conditions | Ambient temperature 20¡À10oC, integration time >200 ms. Data valid for one year after calibration. | ||