| Measurement Accuracy |
| Voltage DC | | 110.000 mV | 0.025%+0.015% * | | 1.10000 V | 0.025%+0.005% * | | 11.0000 V | 0.025%+0.005% * | | 110.000 V | 0.05%+0.005% * | | 300.00 V | 0.05%+0.005% * |
|
| Voltage AC | | 20 to 40 Hz | 2% + 10 | | 40 to 500 Hz | 0.5% + 5 | | 500 to 1 kHz | 2% + 10 | | 1 kHz to 5 kHz | 10% + 20 | | Ranges: | 1.1000, 11.000, 110.00, 300V |
|
| Current DC | | 30.000 mA | 0.01% + 0.015% * | | 110.00 mA | 0.01% + 0.015% * |
|
| Resistance | | 11.000 O | 0.05% + 50 mO | | 110.00 O | 0.05% + 50 mO | | 1.1000 kO | 0.05% + 500 mO | | 11.000 kO | 0.1% + 10 O |
|
| Frequency | | 1.00 to 109.99 Hz | 0.05 Hz | | 110.0 to 1099.9 Hz | 0.5 Hz | | 1.100 to 10.999 kHz | 5 Hz | | 11.00 to 50.00 kHz | 50 Hz |
|
| Pressure | | Accuracy | from 0.025% of range using any of 29 pressure modules. | | (for detailed specifications refer to pressure modules in options and accessories) | | Modules available for differential, gage, vacuum, absolute, dual and high pressure. |
|
| Note | | * | (% of reading + % of full scale) |
|