Product Description

The HP/Agilent 4194A Impedance/Gain-Phase Analyzer is an integrated solution for efficient measurement and analysis or go/no-go testing of components and circuits. Detailed impedance and transmission characteristics including secondary parameter derivations can be simply and quickly evaluated or tested. The HP/Agilent 4194A can contribute to improving engineering productivity and reducing test cost. The analyzer is flexible and has a wide measurement capabilities in both impedance and transmission measurements, plus it is fully programmable using Auto Sequence Programming (ASP). Desired measurements and computations, including graphics analysis, can be programmed simply by storing front panel keystroke operations, allowing you to customize measurement, computation and analysis functions. The HP / Agilent 4194A also features high accuracy and error elimination functions to ensure reliable measurements.

Options:

4194A Impedance/Gain-Phase Analyzer
Opt 350*: 50 Ohm System
Opt 375*: 75 Ohm System
Opt W30: 3-year hardware support
Opt 001: High Stability Frequency Reference
HP 41941A* Impedance Probe Kit (1.5m)
HP 41941B* Impedance Probe Kit (3 m)
*Must select either OPT.350 or 375

Features:

  • High Accuracy and Wide Range:

    • Impedance Measurement: 100Hz to 40MHz, 0.1mΩ to 1.6MΩ,0.17%

    • 10 kHz to 100 MHz, 0.1Ω to 1MΩ, 1.5% when used with the HP / Agilent 41941A/B

    • Gain-Phase Measurement: 10 Hz to 100MHz, -107dBm to +15dBm, 0.1dB resolution


  • Flexible Measurement, Computation and Analysis Capabilities on a Color Graphic Display

  • Fully Programmable


Specifications:

Impedance Measurements
  • Measurement Parameters: |Z|, |Y|, θ, R, X, G, B, L, C, D, Q. 20 parameter combinations are available.

  • Test Frequency: 100 Hz - 40 MHz (cable length: 0m), 100Hz - 15 MHz (cable length: 1m), 1mHz resolution

  • OSC Level: 10 mV - 1Vrms (≤10MHz), 10mV - 0.5Vrms (> 10MHz), 3 digit resolution

  • DC Bias:0 - +40V, 10mV resolution

  • Measurement terminal: 4 - terminal pair configuration

  • Measurement Range and Max. Resolution:

    Measurement
    Parameter
    RangeMax. Resolution
    |Z|, R, X10mΩ to 100MΩ100μΩ
    |Y|, G, B10nS to 100S1nS
    θ+ 180°0.01°
    L1nH to 100kH10pH
    C10fF to 0.1F0.1fF
    D0.001 to 100.0001
    Q0.1 to 10000.1


  • Basic Measurement Accuracy: 0.17%