The Astro-Med Dash 18 field chart recorder is a compact, 18-channel data acquisition recorder ("chart recorder") ideal for medical research applications. The observed waveform data is captured and shown in real-time on a 15.4" color display and stored in memory on an 18 GB hard-drive and optional array recorder.


The Dash 18 is extremely versatile with the capability to capture video and audio data internally. Universal inputs let you connect nearly any signal without using external signal conditioning. The Dash 18 can record a wide range of voltages, temperature, DC Bridge and frequency inputs- all independently set up and enhanced with customizable control buttons.

Specifications:

Color Display
TypeActive matrix color LCD (TFT) with 15.4" viewing area
Resolution1280 x 1024
TouchscreenFull-screen, resistive
FunctionsUser interface with touch-based icons and menus; Real-time waveform monitoring; Review previous waveform records while recording; Overlay numeric values in Engineering Units
Real-time Signal Processing
FiltersLow Pass Stops: 10 Hz to 10,000 Hz; High Pass Starts: 0.1 Hz to 100 Hz; Notch: 50 or 60 Hz center
RMSTime constant selectable from 0.02 to 2 seconds
Frequency to Voltage5 Hz to 20 kHz, ±0.05% of measurement +0.1 Hz
Cross-ChannelSum, Difference, Ratio, Product (Power)
Real-time Signal Testing
BasicAll active signals monitored simultaneously
Test TypesTests include window, slope/level, slew and event pattern
OutputsResult available for alarms, trigger, abort and external TTL
Data Capture
Recording MethodInternal 18 Gbyte hard drive
Maximum Rate100,000 samples/second per channel (all channels)
Minimum Rate1 sample/minute
Dual RateSecond slower rate available when main rate is 50 kHz or less
Total CapacityOver 4 billion samples
Maximum Record2 billion samples
HeaderInformation on units, range, sample rates, etc. saved with data
EventsAll standard event inputs captured with waveforms
Trigger Point LockAmounts of pre- and post-trigger is user adjustable
Auto Re-armAllows automatic stacking of captures
Auto PlaybackYes
External Sample RateTTL sample clock up to 40,000 Hz
Signal Conditioner Specifications (General)
Channels per Module6
Isolation250 Vrms (iso-common to chassis and other iso-commons)
Bandwidth12 kHz (-3 dB)
Input CouplingDC
Zero SuppressionYes
User Engineering UnitsYes
CalibrationSemi-automated to external reference
Single-ended Voltage Input
ConnectorGuarded banana jack
Maximum Rated Input±250 Vrms
Maximum Transient Input±400 volts
Specified Ranges40 to 400 VFS; 4 to 40 VFS; 0.4 to 4 VFS
Accuracy (25° C)±0.5% of full scale
Mininum Input Impedance1 Meg ohms
Differential Voltage Measurements (use for bridge measurements)
Input TypeIsolated, differential
ConnectorScrew terminal header
Absolute Maximum Input±40 volts differential
Measuring Ranges200 to 1600 mVFS; 50 to 500 mVFS; 5 to 50 mVFS4 to 4 VFS
Minimum Input Impedance200k ohms
CMR at 60 Hz>60 dB
Auto BalanceYes, limited to maximum span
Thermocouple Measurements
Input TypeIsolated, differential with screw terminal header connector
Absolute Maximum Input±40 volts
Specified RangeType J: 0 to 760° C/Type K: 0 to 1370° C/Type T: -160 to 400° C/Type E: -100 to 1000° C/Type N: 0 to 1300° C
Accuracy (25° C)0.5% of measurement +2° C
Bandwidth10 Hz (-3 dB)
Intrinsic Noise0.5° C (1 Hz filter)
LinearizationNIST polynomial
Units°C or °F
Frequency Measurements
MethodZero-crossing, channels 1, 7 and 13 only
Signal AmplitudePeak-to-peak minimum is 20% of attenuator
HysteresisYes
Power
Input Voltage Range102 to 264 VAC
Frequency Range47 Hz to 63 Hz
Power Consumption300 watts maximum
Compliance
SafetyEN 61010-1 EMC FCC Class A, EN 61326
Power HarmonicsIEC1000-3-2