Agilent
N9320B RF Spectrum Analyzer
9 kHz to 3.0 GHz
Frequency and Time Specification
Frequency
Supplemental information
Range 9 kHz to 3 GHz AC coupled
100 kHz to 3 GHz Preamp on
Resolution 1 Hz
Internal 10 MHz frequency reference
Aging rate ± 1 ppm/year
Temperature stability ± 1 ppm 5 to +45 °C, reference 25 °C Supply voltage stability ± 0.3 ppm
Residual FM ≤ 100 Hz p-p in 100 ms nominal RBW = 1 kHz, VBW = 1 kHz
Frequency readout accuracy (start, stop, center, marker)
Marker resolution (freq span)/(number of sweep point -1)
Uncertainty ± (freq indication x freq reference uncertainty1 + 1% x span + 20% x resolution bandwidth + marker resolution)
Sweep point 461, fixed
Marker frequency counter
Resolution 1 Hz, 10 Hz, 100 Hz, 1 kHz Selectable
Accuracy ± [(marker freq x freq reference uncertainty1) + (counter resolution)]
Frequency span (FFT and swept mode)
Range 0 Hz (zero span), 100 Hz to 3.0 GHz
Resolution 1 Hz
Accuracy ± span/(swept points -1)
Sweep time and triggering
Span range 10 ms to 1000 s Span > 0 Hz
6 μs to 200 s Span = 0 Hz (minimum resolution = 6 μs) Mode Continuous, single
Trigger Free run, video, external
Trigger slope Positive or negative edge Selectable
Trigger delay 0 to 80 sweep time
Resolution bandwidth (RBW)
Range (-3 dB bandwidth) 10 Hz to 1 MHz, in 1-3-10 sequence
Accuracy ± 5% nominal
Resolution filter shape factor < 5:1 nominal
Range (-6 dB bandwidth) 200 Hz, 9 kHz, 120 kHz, 1 MHz EMI bandwidth (CISPR 16-1-1 complaint), requires Option EMF
Accuracy ± 10% nominal
Resolution filter shape factor < 5:1 nominal -60 dB/-6 dB bandwidth ratio
Video bandwidth (VBW)
Range 1 Hz to 1 MHz in 1-3-10 sequence -3 dB bandwidth
1. Frequency reference uncertainty = Aging rate x period since adjustment + supply voltage stability + temperature stability.
3
Amplitude Specifications
Amplitude range
Measurement range 10 MHz to 3 GHz: Displayed average noise level (DANL) to +30 dBm
(PA OFF) 1 to 10 MHz: DANL up to 23 dBm
100 kHz to 1 MHz: DANL up to 20 dBm
Input attenuator range 0 to 70 dB, in 1 dB steps
Maximum damage level
Supplemental information
Average continuous power ≤ +37 dBm Input attenuator setting ≥ 10 dB, 3 minutes maximum
Peak pulse power ≤ +50 dBm (100 W) For < 10 μs pulse width, < 1% duty cycle, and input attenuation ≥ 40 dB
DC voltage 50 VDC maximum
Level display range
Log scale units dBm, dBmV, dBμV, dBμA
Linear scale units μV, mV, V, μA, mA, A, μW, mW, W
Marker level readout 0.01 dB Log scale Resolution 0.01% of reference level Linear scale Number of traces 4
Detectors Positive-peak, negative-peak, sample, normal, RMS
Trace function Clear/write, maximum hold, average, minimum hold, view
Frequency response
10 dB input attenuation, reference: 50 MHz, 20 to -30 °C
200 kHz to 2.0 GHz ± 0.5 dB Preamp off
2.0 to 3.0 GHz ± 0.7 dB Preamp off
1 MHz to 2.0 GHz ± 0.6 dB Preamp on
2.0 to 3.0 GHz ± 0.8 dB Preamp on
Input attenuation switching uncertainty at 50 MHz
Attenuation > 2 dB, preamp off
0 to 60 dB attenuation ± 0.4 dB Relative to 10 dB (reference setting)
Absolute amplitude accuracy
Center frequency 50 MHz, RBW 1 kHz, VBW 1 kHz, amplitude scale log, span 100 kHz, sweep time coupled, peak detector, signal at reference level
Preamp off ± 0.3 dB Reference level -10 dB, input attenuation
10 dB
Preamp on ± 0.4 dB Reference level -30 dB, input attenuation
10 dB
4
Amplitude Specifications (continued)
Level measurement uncertainty
Supplemental information
20 to -30 °C; frequency > 1 MHz; signal input 0 to -50 dBm; reference level 0 to -50 dBm; input attenuation 20 dB; RBW
1 kHz, VBW 1 kHz; after calibration; preamp off
Overall amplitude accuracy ± 1.5 dB
± 0.5 dB, typical
Level display range
Log scale units dBm, dBmV, dBμV, dBμA
Linear scale and units W, mW, μW, A, mA, μA, V, mV, μV Marker level readout 0.01 dB
Resolution 0.01% of reference level Log scale Number of traces 4 Linear scale Detectors Positive-peak, negative-peak, sample,
normal, RMS
Trace functions Clear/write, maximum hold, average, minimum hold, view
Preamplifier
Frequency range 1 MHz to 3.0 GHz
Gain 18 dB nominal
5
Dynamic Range Specifications
1 dB gain compression
Supplemental information
Preamp off 50 MHz to 3.0 GHz > 0 dBm, typical; total power at input mixer
Preamp on 50 MHz to 3.0 GHz > -20 dBm, typical; total power at the preamp
Total power at the preamp = total power at the input
(dBm) - input attenuation (dB)
Displayed average noise level (DANL)
Input terminated, 0 dB RF attenuation, RBW = 10 Hz, VBW = 1 Hz, sample detector
Preamp off 9 to 100 kHz < -90 dBm, nominal
100 kHz to 1 MHz < -90 dBm – 3 x (f/100 kHz) dB
1 to 10 MHz < -124 dBm
10 MHz to 3 GHz < -130 dBm + 3 x (f/1 GHz) dB Preamp on 100 kHz to 1 MHz < -108 dBm – 3 x (f/100 kHz) dB
1 to 10 MHz < -142 dBm
10 MHz to 3 GHz < -148 dBm + 3 x (f/1 GHz) dB
Spurious response
Preamp off, signal input -30 dBm, 0 dB RF attenuation
Second harmonic distortion 10 to 200 MHz +30 dBm
200 to 500 MHz +35 dBm
500 MHz to 3 GHz +43 dBm
Preamp off, signal input -30 dBm, 0 dB RF attenuation
Third-order intermodulation (TOI) 300 MHz to 3 GHz +10 dBm; +13 dBm nominal
6
Dynamic Range Specifications (continued)
Spurious response (continued)
Supplemental information
Input related spurious < -60 dBc -30 dBm signal at input mixer, 20 to 30 °C
Residual response (inherent) < -83 dBc Input terminated and 0 dB RF attenuation, preamp off
Phase noise Specification Typical
Offset from CW signal 10 kHz < -88 dBc/Hz < -90 dBc/Hz
Fc = 1 GHz, RBW = 1 kHz, VBW = 10 Hz, and sample detector, log average, average times > 40
100 kHz < -100 dBc/Hz < -102 dBc/Hz
1 MHz < -110 dBc/Hz < -112 dBc/Hz
Residual FM ≤ 100 Hz peak-to-peak in 100 ms
1 kHz RBW, 1 kHz VBW
Tracking Generator Specifications (Option TG3 required)
Output frequency
Supplemental information
Range 100 kHz to 3 GHz Settable to 9 kHz
Resolution 1 Hz
Output power level
Range -30 to 0 dBm
Resolution 0.1 dB
Absolute accuracy ± 0.75 dB 20 to 30 °C, at 50 MHz with coupled source attenuator, referenced to -20 dBm
Output flatness ± 3 dB 100 kHz to 10 MHz
± 2 dB 10 MHz to 3 GHz
VSWR < 1.5:1 300 kHz to 3 GHz, input attenuator ≥ 12 dB Connector and impedance N-type female, 50 Ω
Maximum safe reverse level
Average total power 30 dBm (1 W) AC coupled 0 VDC MAX
7
Modulation Analysis Specifications
Demodulation
Frequency range 10 MHz to 3 GHz
Supplemental information
Carrier power accuracy ± 2 dB ± 1 dB typical Input power -30 to +20 dBm Auto attenuation Carrier power displayed resolution 0.01 dBm
AM measurement (included in Option AMA)
Modulation rate 20 Hz to 100 kHz
Accuracy 1 Hz, nominal Modulation rate < 1 kHz
< 0.1% modulation rate, nominal Modulation rate ≥ 1 kHz
Depth 5 to 95% Accuracy ± 4% nominal FM measurement (included in Option AMA) Modulation rate 20 Hz to 200 kHz
Accuracy 1 Hz, nominal Modulation rate < 1 kHz
< 0.1% modulation rate, nominal Modulation rate ≥ 1 kHz
Deviation 20 Hz to 400 kHz Accuracy ± 4% nominal ASK measurement (included in Option DMA)
Symbol rate range 200 Hz to 100 kHz
Modulation depth/index range 10 to 90%
Accuracy ± 4% of reading, nominal
Displayed resolution 0.1% FSK measurement (included in Option DMA) Symbol rate range 1 to 100 kH FSK deviation range 1 to 400 kHz
Accuracy ± 4% nominal β ≥1 and β≤4, β is the ratio of frequency deviation to symbol rate
Displayed resolution 0.01 Hz
8
Inputs and Outputs
Front panel
RF input connector N-type female, 50 Ω
Supplemental information
VSWR < 1.5:1 300 kHz to 3 GHz, input attenuator ≥ 10 dB Calibration output Amplitude -10 dBm ± 0.3 dB
Frequency 50 MHz
Accuracy Same as the frequency reference
Connector and impedance BNC-type female, 50 Ω Probe power Voltage/current +15 V, 150 mA maximum
-12.6 V, 150 mA maximum
RF output connector N-type female, 50 Ω Option TG3 installed
USB interface (host) A plug, version 1.1
Rear panel
10 MHz reference output Output amplitude > 0 dBm
Connector and impedance BNC-type female, 50 Ω
10 MHz reference input Input amplitude -5 to +10 dBm
Frequency lock range ± 5 ppm of specified external reference input frequency
Connector and impedance BNC-type female, 50 Ω External trigger input Input amplitude 5 V TTL level
Connector and impedance BNC-type female, 10 kΩ
VGA output VGA analog RGB 31.5 kHz horizontal, 60 Hz vertical sync rates, non-interlaced
D-sub 15-pin female connector VGA compatible
640 x 480 screen resolution LAN TCP/IP interface 10Base, RJ-45 connector USB interface (device) B plug, version 1.1
GPIB interface IEEE-488 bus connector Optional G01 installed
9
General
Temperature range
Operating +5 to +45 °C Storage -20 to +70 °C EMC
Complies with European EMC Directive 2004/108/EC
• IEC/EN 61326-1 or IEC/EN 61326-2-1
• CISPR Pub 11 group 1, class A
• AS/NZS CISPR 11:2004
• ICES/NMB-001:2004
This ISM device complies with Canadian ICES-001
Safety
Complies with European Low Voltage Directive 2006/95/EC
• IEC/EN 61010-1 2nd Edition
• Canada: CSA C22.2 No. 61010-1-04
• USA: UL 61010-1 2nd Edition
Audio noise
Acoustic noise emission
LpA < 70 dB Operator position Normal position Per ISO 7779
Environmental stress
Supplemental information
Samples of this product have been type tested in accordance with the Agilent Environmental Test Maunal and verified to be robust against the environmental stresses of storage, transportation, and end-use; those stresses include, but are not limited to, temperature, humidity, shock, vibration, altitude, and power line conditions. Test methods are aligned with IEC 60068-2
and levels are similar to MILPRF-28800F Class 3
Power requirements
Voltage and frequency (nominal) 100 to 240 VAC, 50 to 60 Hz Auto ranging
Power consumption < 65 W
Display
Resolution 640 x 480
Size 165.1 mm (6.5 in) diagonal (nominal)
Data storage
Internal 16 MB nominal
External Supports USB 2.0 compatible memory devices
10
General (continued)
Weight (without options)
Net 8.4 kg (18 lbs) nominal
Shipping 14.5 kg (32 lbs) nominal
Dimensions
Supplemental information
Height 132.5 mm (5.2 in) 3U rack height
Width 320 mm (12.6 in) Length 400 mm (15.7 in) Warranty
The N9320B spectrum analyzer is supplied with a one-year warranty
Calibration cycle
The recommended calibration cycle is one year. Calibration services are available through Agilent Service Centers