Bio-Rad Benchmark Microplate Reader
This is a Used Bio-Rad Benchmark Microplate Reader with Six Filters Installed (340, 405, 415, 490, 595, 655nm). Unit Has Been Tested Against a Calibration Plate and the Results are Provided Below. Unit is in Good Working Condition and Physically it Has Some General Wear, Nicks-Scratches, and Some Exterior Casing Yellowing. Features, Shaking, Incubation, Dual/Single Wavelength Reading, and On Board Printer. For Complete Specs, Please Download the User's Manual Here: Manual. Unit Will Guaranteed for 30 Days Allowing Enough Time for Your Lab To Fully Test and Inspect. Unit Comes as Pictured with Power Cord, User's Manual, and Two Rolls of Printer Paper (Please Note: No External Data Management Software is Included: Raw Data Is Obtained From the On Board Printer). Please Add $45 Flat Rate Ground Shipping.
This Reader's Calibration Test Plate Results:
1) 340nm Reading: * No Calibration Filter Available
2) 405nm Reading: * Expected .595, 1.250, 1.852 * Actual: .599, 1.269, 1.883
3) 415nm Reading: * No Calibration Filter Available
4) 490nm Reading: * Expected .545, 1.150, 1.705 * Actual: .553, 1.164, 1.722
5) 595nm Reading: * Expected .555, 1.180, 1.755 * Actual: .564, 1.192, 1.764
6) 655nm Reading: * No Calibration Filter Available
Specifications:
- Spectral range: 340-700 nm
- Filter Wheel Capacity 6: Five Included: (340, 405, 415, 490, 595, 655nm)
- Photometric Range: 0.000 - 3.500 O.D.
- Read times: 25 seconds single wavelength
- Mixing/Shaking: Yes
- Compatible Plates: Polystyrene flat, U- and V-bottom, PVC flex plates
- Analysis Options: Raw data, absorbance, matrix, limit, cutoff, concentration
- Printer: 20 column, 5 x 7 dot-matrix printer built-in
- Computer interface: Bi-directional RS232 serial communication port
- Resolution: 0.001 O.D.
- Accuracy: +/- 1.5%
- Reproducibility: Less than 2.0%
- Stability and drift: Automatic calibration before each plate reading
If This Unit Does Not Meet Your Needs, Please See: Other Microplate Reader Available