PORTABLE VENTILATOR SPECIFICATIONS by Nanjing Foinoe Co., Ltd
- Item Condition:
- Brand:
- Model#:
- Location:
- Price:
Product Description
Used for first aid, ambulance andpatients transport within hospital,
it can be used for pediatric and adult patients.
Physical Specifications
Ventilator Dimensions
Width: 300mm Depth:140mm Height:230mm
Carrying packages Dimensions
Width: 460mm Depth: 160mm Height: 350mm
PackagingDimensions
Width: 620mm Depth:410mm Height:240mm
VentilatorWeight:3Kg
Carry bag Weight (with cylinder ): 7Kg
PackagingWeight: 11Kg
Environmental Specifications
Operating Requirements
Temperature: 5 to 40 ¡æ;
Relative Humidity: less than 80%;
The atmosphere pressure: 70 to 106 kPa ;
Storage/Shipping Requirements
Temperature: -20 to 70 ¡æ
Relative Humidity: less than 93%
The atmosphere pressure: 70 to 106 kPa
Electrical Specifications
Internal battery DC12V 2200mA
Input Power Supply Requirements
DC 12V 10 VA;
Charger Power Supply Requirements
AC 220V+/-15% 50HZ+/-2% 25 VA;
Pneumatic Specifications
Oxygen Inlet
Pressurized Oxygen: 0.4 MPa ¡¨¤10% medical oxygen.
(Availability through Oxygen Pressure Reducerprovided as a accessory)
What Is Provided
Parts Name | Number | Specification |
Ventilator | 1 |
|
Oxygen Cylinder | 1 | 2.5Lcan be used for 1 hour |
Carry bag | 1 | With cylinder |
Breath Circuit | 1 | 1.2m |
Breath Value | 1 |
|
Mask | 1 |
|
Chargers | 1 |
|
Battery | 1 | Can be used for 8 hours |
Wrench Tool | 1 |
|
Oxygen Pressure Reducer | 1 | For NormalOxygen Cylinder |
OxygenBridge | 1 |
|
Hand Book | 1 |
|
CD For Operation | 1 |
|
Performance Specification And Tolerances
Ventilator Parameters | Ranges | Remark |
Controls | ||
Power |
| ON/OFF |
Modes CMV(IPPV) A/C(SIPPV) SIMV SPONT PEEP(Choice for suppliers) | ||
Tidal Volume | 50~1200ml |
|
Breath Rate | 5~60 bpm 1~12 bpm for SIMV |
|
Sigh | 1~10 per 100 breath |
|
%O2 | 45~90% |
|
Trigger Level | -10~0hpa |
|
PIP | 0~60hpa |
|
Base Flow | 0~5L/min |
|
Peep | 0~10 hpa | (Choice of suppliers) |
Over Pressure Relief | < 60hpa |
|
Performance Specification And Tolerances
Ventilator Parameters | Ranges | Tolerances |
Alarms and Indicators | ||
High Pressure Alarm Limit | 20~60hpa to be set flash, red indicator | +/-0.1kpa |
Low Pressure Alarm Limit | 2~20hpa to be set Red indicator | +/-0.1kpa |
Low Battery Voltage | Surplus energy can work for an hour red indicator |
|
Alarm | Red indicator |
|
Alarm Silence | Yellow indicator | 120s |
Power | Green indicator |
|
Charging Indicator | Green indicator |
|
Performance Specification And Tolerances
Ventilator Parameters | Ranges | Remark |
Monitors | ||
Main Circuit Pressure | -20~100hPa |
|
Wave Display | ||
Main Circuit Pressure | ||