The Lmax Microplate Luminometer is designed for bio- and chemiluminescence measurements using 96-well and 384-well microplates. Combining high sensitivity with extremely low crosstalk and programmable reagent injection, the Lmax is suitable for glow luminescence measurements in both 96- and 384- well plates. By making use of up to two injectors, it is also suitable for measuring flash luminescence in 96-well plates. The Lmax uses a special version of SOFTmax PRO for Lmax software from Molecular Devices which provides complete instrument control and powerful data processing capability. Multi-phase measurements can be run as required (for example, in ATP measurements for sterility tests), taking into account an extraction time after the first injection.
The Lmax is well suited for the following applications:
- Luminescence-based immunoassays, including glow luminescence with dioxetane.
- EIAs based on alkaline phosphatase or peroxidase activity.
- DNA hybridization assays for the detection of pathogenic microorganisms and viruses.
- All applications of the (coupled) ATP measurement via the luciferinluciferase system.
- All applications of the (coupled) NAD(P)H measurement via bacterial luciferases.
- Reporter gene assays via the reporter gene: firefly luciferase, aquorin, Beta-galactosidase (lacZ), alkaline phosphatase or glucuronidase. - Cellular luminescence (phagocytosis assay).
The "Ultra-Fast Photon Counting" technology utilized in the Lmax, in conjunction with a special configuration of the photomultiplier reading head, allows the Lmax to achieve a reading sensitivity that matches the sensitivity of traditional tube luminometers. Dependent only on the background caused by the reagent, the instrument is capable of detecting ATP quantities of less than 0.01 pg. The dynamic range covers 6 powers of ten. An optical filter may be inserted between sample and photomultiplier to filter out specific wavelength ranges.
SPECIFICATIONS:
Microplate formats: 96- and 384-well standard microplate (not transparent) in compact or strip-rack format
Detector: Ultra-fast photon counter with overload detection; spectral range 380-630 nm with a dark count rate of less than 250 cps
at 20¡ãC ambient temperature
Sensitivity: Higher than 20 attomole ATP
Crosstalk: Less than 0.01%
Read time: 0.1 up to 1000 sec in steps of 0.1 sec
Dynamic range: More than 6 decades
Temperature control (chamber): Ambient +5¡ãC to 45¡ãC in steps of 0.1¡ãC; optional, three heating zones, microprocessor controlled
Injectors Two: one in the reading head and the other directed to the preceding well
Injection volumes: Freely selectable between 25 and 300 ¨¬L for injectors P ("preinjection") and M (reading or "measurement" position); precision ¡À3¦ÌL
Delay: Delay between injection in reading position and measurement: adjustable from 0 to 1000 sec; between injection before the reading position and further steps: 1.6 to 1000 sec for P-injector; 0 to 1000 sec for M-injector; each adjustable in steps of 0.1 sec
Injection/Read wells: Rectangular from 1 to 96 or 1 to 384 wells. For 96-well plates, any well or combination of wells may be designated for one, two, or no injections (injection cannot be used with 384-well plates).
Ports: Serial RS232 port for computer connection
Waste disposal: Automatic to waste bottle