Sorry, the search criteria can not be empty
Search
View All Categories

Add to Inquiry Cart



<<    <       1       >    >>      GO

Item added to Inquiry Cart

Continue shopping Go to Inquiry Cart
  • Product Overview

The HP 8754A network analyzer consists of an internally leveled, 4 to 1300 MHz swept source. Polar or rectangular measurements are displayed on a CRT through two independent channels. With appropriate signal routing accessories, the 8754A is a fully integrated stimulus/response test system that measures magnitude and phase characteristics of linear networks by comparing the incident signal with the signal transmitted by the device under test or reflected from its input.

The receiver provides 80 dB dynamic range in two independent measurement channels to provide simultaneous measurement of any two transmission or reflection parameters using a single test setup. The basic transmission measurements are insertion loss or gain, insertion phase, and transmission coefficient (S21 or S12). The basic reflection measurements are return loss, from which SWR can be calculated, and reflection coefficient (S11 or S22), from which impedance can be calculated or read from a Smith chart overlay.

A full line of 50 ohm and 75 ohm test sets let you tailor your test setup for specific measurements with the minimum of equipment, or provide maximum versatility for a wide range of applications.

  • Frequency range: 4 MHz to 1300 MHz
  • Sweep modes: Linear full sweep and calibrated sweep widths with variable start or center frequency
  • Markers
    • Internal, crystal-generated harmonic markers; amplitude markers for rectangular displays, intensity markers for polar displays
    • Spacing: 1, 10, and 50 MHz
    • Accuracy: ±0.01%
  • Output power: Calibrated 0 to +10 dBm, accuracy ±0.8 dB, flatness ±0.5 dB
  • Three input channels, R, A, B
    • Two test inputs, A, B with dynamic range of 80 dB
    • Reference input, R, with dynamic range of 40 dB, minimum input level of -40 dBm
  • Maximum input level: 0 dBm
  • Damage level: +20 dBm (50 Vdc)
  • Noise level: <-80 dBm
  • Crosstalk between channels: >83 dB
  • Receiver flatness
    • Absolute (A, B, R): ±1 dB
    • Ratio (A/R, B/R): ±0.3 dB
  • Phase range: ±180°
  • Phase frequency response: ≤±2.5°
  • Display: CRT displays either polar trace or two independent rectangular traces
    • Channel 1: A Magnitude Absolute (dBm), R Magnitude Absolute (dBm), A/R Magnitude Ratio (dB), B/R Magnitude Ratio (dB)
    • Channel 2: B Magnitude Absolute (dBm), B/R Magnitude Ratio (dB), B/R Phase (degrees)
    • Polar: A/R Magnitude Ratio (dB) and Phase (degrees) 
  • External sweep input: 0 to +10 V nominal
  • Sweep output: -5 V to +5 V nominal
  • External marker input: Typically, -13 dBm RF signal into the External Marker Input will produce an amplitude or intensity marker on the trace at the frequency of the RF signal
Status Discontinued
Alternative E5061A ENA-L RF Network Analyzer