  • Product Overview

Bandwidth : DC to 3.5 GHz
Risetime : 100 ps
Attenuation : 10X
Input Resistance : 500O
Input Capacitance : approx. 1 pF
Scope Input Resis : 50O
Readout Pin : Yes
Ground Ref : No
ID Button : Yes
Max Rated Voltage : 16 V RMS

COMPATIBILITY

Catalogs specifically mention the
7A19, 7A24, 7A29
485, DC508A(P6056 only), SI5010(P6057), DC5010(P6056), 11000 series

CHARACTISTICS

   The P6056 (10X), and P6057 (100X), are miniature, fast-rise, low input-capacitance probes with relatively constant input resistance over a wide range of frequencies, designed for use with 50 O wide-band oscilloscopes such as the Tektronix 7904 with 7A19 plug-in installed.

   Cable length for each probe is selected for calibrated signal delay.

   These probes can also be used with sampling systems by utilizing appropriate adapters.

   Both probes are equipped with a special BNC connector that provides trace identify and digital readout information when used with plug-in units or mainframes that have these features. A button on the Trace Shift Sleeve portion of the probe body provides the trace identify function.


   above excerpt from Tektronix P6056/P6057 Probe Manual

SPECIFICATIONS

       Electrical characteristics are applicable when used with an instrument that is calibrated within an ambient temperature between +20°C and +30°C. Unless otherwise stated, electrical characteristics apply over and operating temperature range from -15°C to +55°C, and to an altitude of 15,000 feet.

       NOTE

       The following electrical specifications apply when the probe is operating into a 50 O oscilloscope input or a 50 O termination.
       ELECTRICAL

       Attenuation: (Includes ±2% Input Resistance of Oscilloscope)    P6056, 10X within 3%
       P6057, 100X within 3%.    
       Input Resistance: (Within 3%)    P6056, 500 O DC to 300 MHz, decreasing to 250 O at 1 GHz.

       P6057, 5000 O DC to 300 MHz, decreasing to 1500 O at 1 GHz

           
       Input Capacitance:    P6056 and P6057: Approximately 1 pF.

       Bandwidth (-3 dB): (Calculated from Risetime)    P6056: DC to 3.5 GHz (probe only), DC to approximately 500 MHz with 7A19 and 7904
       P6057: DC to 1.4 GHz (probe only), DC to approximately 480 MHz with 7A19 and 7904.

       
   Risetime:    P6056: 100 ps (probe only).
   P6057: 250 ps (probe only).

   Maximum Input Voltage: (Limited by 0.5 W nose resistor dissipation, and derated with frequency).    P6056: 16 V DC or RMS (AC or pulse) to 800 MHz, decreasing to 9 V at 1 GHz..
   P6057: 50 V DC or RMS (AC or pulse) to 500 MHz, decreasing to 21 V at 1 GHz.

   Signal Delay Time:    P6056 and P6057: 8.2 ns (6 ft probe), 12.3 ns (9 ft probe), each within 35 ps.
   GENERAL

   Weight:    6 ft probe - 2 ounces, 9 ft. probe - 3 ounces (complete with accessories, except Instruction Manual). Shipping: 6 ft probe - 12 ounces, 9 ft probe - 13 ounces.
   Dimensions:    Nominally 6 ft. and 9 ft. long (Calibrated for signal delay time).
   Environmental:    Probe will operate within specifications over the following ranges:

   Temperature - From -15°C to +55°C.
   Altitude - To 15,000 feet..
   above excerpt from Tektronix P6056/P6057 Probe Manual